Device Interface - Resonance R17A
Covered by this topic
This interface uses the Resonance cloud interface to import results from the audiometer to Enterprise Health .
Import Instructions
- Create a new patient on the resonance audiometer. If the system uses multiple partitions, the patient ID should be the partition identifier and MRN without a dash character in between. For example, William Hart would be TEST10019. If only one partition is used, the partition identifier can be excluded.
- Perform a test on the Resonance audiometer. The results will be automatically sent to Enterprise Health . On success, the test will be removed from the audiometer.
- Once results have been received, they can be viewed in the patient’s chart. As this device does not support location information, the location on the audiogram document created in Enterprise Health will be blank.
Collected Fields
The following fields are collected:
- Left ear frequencies: 500Hz, 1kHz, 2kHz, 3kHz, 4kHz, 6kHz, 8kHz
- Right ear frequencies: 500Hz, 1kHz, 2kHz, 3kHz, 4kHz, 6kHz, 8kHz
- Patient MR Number
- Test Date/Time
- Audiometer Serial Number
- Calibration Date/Time
Installation Help
- Resonance will need to create a new Endpoint in their Resodis software to send data from devices to Enterprise Health .
- The only authentication that can be used from Resonance to Enterprise Health is HTTP Basic Authentication. A new user will need to be created with a password. Once that is done, the username and password will need to be base64 encoded and entered into the Resodis Endpoint manager.
- The installation of this interface requires additional help from the Enterprise Health team. Please contact your deployment or support representative.
Enterprise Health Documentation